DE-3506 AC Clamp Meter
Data/Range hold ;Clamp Size ∅40mm
4000 counts
- Measurement Functions:ACA, ACV, DCV, Resistance, Capacitance, Continuity, Frequency, Duty, Diode, REL.
- Clamp Size:φ40mm or 10x47mm
- LCD Display:3 3/4 Digits LCD With 4000 Counts
- Applicable Circuit Voltage:AC 600V or less
- Sampling Reading Rate:3 Times/ Sec. Approx.
- Range:Automatic & Manual For Selection
- Applicable Height:Up to 2000m (above sea level)
- Auto Power Off:Approx. 30 Minutes
-
Low Battery:
Display on LCD
- Battery: AA 1.5V x2
- Safety standard:IEC 61010-1 , IEC 61010-2-032 600V CAT III Pollution Degree 2
Measurement Ranges:
(23°C ±5°C, 80% R.H. MAX) Accuracy:±(%rdg+dgt)
Measurement |
Range |
Resolution |
AVG Accuracy |
Input Impedance |
Remarks |
|
DE-3506 |
DE-3507 |
|||||
(ACA) |
400A |
0.1A |
1.5 %+10 |
──── |
|
|
1000A |
1A |
|||||
(ACV) |
400mV |
0.1mV |
1.5%+10 |
Approx.≥ 11MΩ |
|
|
4V |
0.001V |
Approx. 11MΩ |
|
|||
40V |
0.01V |
Approx. 10MΩ |
||||
400V |
0.1V |
1.5%+5 |
||||
600V |
1V |
|||||
(DCV) |
4V |
0.001V |
0.7%+3 |
Approx. 11MΩ |
|
|
40V |
0.01V |
1%+3 |
Approx. 10MΩ |
|||
400V |
0.1V |
|||||
600V |
1V |
|||||
(Resistance) |
400Ω |
0.1Ω |
1%+5 |
──── |
*Open voltage:approx.0.4V *The measuring current changes in accordance with the resistance measured. |
|
4kΩ |
0.001kΩ |
|||||
40kΩ |
0.01kΩ |
|||||
400kΩ |
0.1kΩ |
|||||
4MΩ |
0.001MΩ |
3%+5 |
||||
40MΩ |
0.01MΩ |
5%+5 |
||||
(BUZZER)) |
400Ω |
0.1Ω |
The buzzer turns on for resistances below approx.100Ω |
──── |
|
|
(DIODE) |
1.000V |
0.001V |
10%+5 |
──── |
Open voltage: Approx.1.5V |
|
(CAP) |
50nF |
0.01nF |
3%+10 |
──── |
Accuracy is subject to zero set by pressing REL. key before forwarding measurement. |
|
500nF |
0.1nF |
|||||
5μF |
0.001μF |
|||||
50μF |
0.01μF |
5%+15 |
||||
100μF |
0.1μF |
|||||
Hz |
5.000Hz |
0.001Hz |
0.7%+5 |
──── |
*Accuracy in the case of sine wave. *5.000Hz~100kHz:typical above 5V rms |
|
50.00Hz |
0.01Hz |
|||||
500.0Hz |
0.1Hz |
|||||
5.000kHz |
0.001kHz |
|||||
50.00kHz |
0.01kHz |
|||||
100.0kHz |
0.1kHz |
|||||
DUTY |
0.1~99% |
0.1% |
──── |
──── |
*About input sensitivity and frequency characteristic: (Square wave DUTY 50% input) 2.5V 0 to peak input:≧1kHz 6V 0 to peak input:≧10kHz 40V 0 to peak input:≧70kHz |
※Remarks:The accuracy data is subject to the conductor to be measured at the center of the iron core (clamp jaw)
TEMP./Humidity:
- Operation:0°C~ 50°C(32℉~122℉)/<80% R.H.(no condensation)
- Storage:-10°C~ 60°C(14℉~140℉)/<70%R.H.(no condensation)
Dimension & Weight:
- 241(L)x62(W,Max91)x34(H,Max45)mm
- Approx.390g (Including battery)
Accessories:
- Battery AA 1.5V x2
- Instruction Manual x1
- Test Lead (black+red) x1
- Carrying case x1
Optional Accessories :
-
AC Line Splitter(DE-10ALS)
*Applicable to electrical appliances small current
- Data Hold
- Capacitance Measurement
- Hz/DUTY Measurement
- Relative Measurement
- Range Hold
- Auto Power Off Selection
- Auto Range & Manual Range
- High Accuracy