DE-208F Digital Multimeter (D.M.M)
Resistance, Diode, Buzzer, Capacitance, Frequency, DUTY, REL.
6600 Counts
- LCD Display:6600 Counts
- Sampling Rate:Approx. 3 Times/Sec.
- Range:Automatic & Manual
- Overload:”OL” Display On LCD
-
Low Battery:
Display On LCD
- Auto Power Off:In 30 Minutes Approx.
- Battery:9V x1
- Safety standard:IEC 61010-1 CAT IV 600V IEC 61010-1 CAT.III 1000V
Measurement Ranges:
23°C ± 5°C max. 80%R.H. Accuracy: ±(%rdg +dgt)
Measurement |
Range |
Resolution |
Accuracy / Responding Hz (sine wave) |
Input Resistance |
Max. Input Voltage |
|
DE-208E |
DE-208F/208G |
|||||
0.5%+10 (50Hz~60Hz) |
||||||
ACV |
660.0mV (manual range) |
0.1mV |
1.0%+10 (40Hz~150Hz) |
1.0%+10 (40Hz~150Hz) |
≧100MΩ |
1000V rms |
6.600V |
0.001V |
1.0%+10 (40Hz~1kHz) |
1.0%+10 (40Hz~1kHz) |
Approx. 11MΩ |
||
66.00V |
0.01V |
1.0%+10 (40Hz~5kHz)
2.0%+10 (5kHz~10kHz) |
1.0%+10 (40Hz~3kHz) |
Approx. 10MΩ |
||
660.0V |
0.1V |
|||||
1000V |
1V |
Measurement |
Range |
Resolution |
Accuracy |
Input Resistance |
Max. Input Voltage |
DCV |
660.0mV (manual range) |
0.1mV |
0.2%+5 |
≧100MΩ |
1000V rms |
6.600V |
0.001V |
Approx. 11MΩ |
|||
66.00V |
0.01V |
Approx. 10MΩ |
|||
660.0V |
0.1V |
||||
1000V |
1V |
Measurement |
Range |
Resolution |
Accuracy |
Max. Input Voltage |
Ω Resistance |
660.0Ω |
0.1Ω |
0.5%+5 |
1000V rms |
6.600kΩ |
0.001kΩ |
|||
66.00kΩ |
0.01kΩ |
|||
660.0kΩ |
0.1kΩ |
|||
6.600MΩ |
0.001MΩ |
1.0%+5 |
||
66.00MΩ |
0.01MΩ |
2.0%+5 |
Measurement |
Range |
Resolution |
Remarks |
Open-loop Voltage |
Max. Input Voltage |
Continuity |
660.0Ω |
0.1Ω |
Buzzer emits at around ≦30Ω |
Approx. -3.3V |
1000V rms |
Measurement |
Range |
Resolution |
Accuracy |
Open-loop Voltage |
Max. Input Voltage |
Diode |
2.000V |
0.001V |
1.0%+2 |
Approx.3.2V |
1000V rms |
Measurement |
Range |
Resolution |
Accuracy |
Remarks |
Max. Input Voltage |
Capacitance |
6.600nF |
0.001nF |
1.5%+10 |
Press “REL” for offset if some reading at 6.6nF/66nF ranges before measurement |
1000V rms |
66.00nF |
0.01nF |
||||
660.0nF |
0.1nF |
||||
6.600μF |
0.001μF |
||||
66.00μF |
0.01μF |
2.0%+10 |
|||
660.0μF |
0.1μF |
||||
6.600mF |
0.001mF |
||||
66.00mF |
0.01mF |
Measurement |
Range |
Resolution |
Accuracy |
Remarks |
Max. Input Voltage |
Hz Frequency |
66.00Hz |
0.01Hz |
0.1%+3 |
Duty: 1%~98.9% (<10kHz3Vp-p) resolution: 0.01%. Out of rating range display 0.00%(<0.9%) or OL (>98.9%) |
1000V rms |
660.0Hz |
0.1Hz |
||||
6.600kHz |
0.001kHz |
||||
66.00kHz |
0.01kHz |
||||
660.0kHz |
0.1kHz |
||||
6.600MHz |
0.001MHz |
||||
66.00MHz |
0.01MHz |
Measurement |
Range |
Resolution |
Accuracy / Responding Hz (sine wave) |
Remarks |
||
40Hz~5kHz |
5kHz~10kHz |
|||||
AC |
μA |
660.0μA |
0.1μA |
0.5%+10 |
1.2%+10 |
Protected by a 630mA / 1000V fuse |
μA |
6600μA |
1μA |
||||
mA |
66.00mA |
0.01mA |
1.0%+10 |
|||
mA |
660.0mA |
0.1mA |
||||
A |
10A |
0.01A |
1.0%+10 |
1.5%+10 |
Protected by a 10A / 1000V fuse |
*11 ~ 20A stands with 30 sec. max.
Measurement |
Range |
Resolution |
Accuracy |
Remarks |
|
DC |
μA |
660.0μA |
0.1μA |
0.5%+5 |
Protected by a 630mA / 1000V fuse |
μA |
6600μA |
1μA |
|||
mA |
66.00mA |
0.01mA |
|||
mA |
660.0mA |
0.1mA |
|||
A |
10A |
0.01A |
1.0%+10 |
Protected by a 10A / 1000V fuse |
*11 ~ 20A stands with 30 sec. max.
ACA & ACV accuracy in Peak mode:
Measurement |
Range |
Accuracy |
Remarks |
ACV |
6.600V |
2.0%+40 |
AC signals are valid ≧ 100 counts |
66.00V |
|||
660.0V |
|||
1000V |
|||
ACuA |
660.0μA |
||
6600μA |
|||
ACmA |
66.00mA |
||
660.0mA |
|||
ACA |
10A |
★Temperature coefficient:
0.1 X (specified accuracy) / °C (<18 °C or >28 °C)
TEMP./Humidity:
- Operation:0°C~ 50°C(32℉~122℉)/<80% R.H.(no condensation)
- Storage:-10°C~ 60°C(14℉~140℉)/<70%R.H.(no condensation)
Dimension & Weight:
- 188(L)x95(W)x52(H)mm
- Approx.405g (excluding battery)
Accessories:
- 9V Battery x1
- Instruction Manual x1
- Test Leads (black + red) x1
- Data Hold
- Peak Hold
- Max/Min Hold
- Auto power off selection
- Auto range & manual range
- Large LCD with dual displays
- Overload protection with audible alarm
-
Backlight:Blue
- Relative check