  • LCD Display:Dual 19999/ 1999 display、
  • Measuring rate:1.2 / second, nominal
  • Range:Automatic
  • Overload:”OL” Display On LCD
  • Low BatteryDisplay On LCD
  • Auto Power Off:In 5 Minutes Approx.
  • Battery:9V x1
  • Safety standard:IEC 61326-1 & IEC 61326-2-2


  Measurement Ranges:

23°C ±5°Cmax. 80%R.H.

Resistance(Parallel / Series mode):

Range

Resolution

100/120Hz

1kHz

10kHz

100kHz

20.000Ω

0.001Ω

1.0%+3*

1.0%+3*

2.0%+3*

200.00Ω

0.01Ω

1.0%+3

0.3%+2

0.3%+2

0.6%+3

2.0000kΩ

0.0001kΩ

0.3%+2

0.3%+2

0.3%+2

0.6%+3

20.000kΩ

0.001kΩ

0.3%+2

0.3%+2

0.3%+2

0.6%+3

200.00kΩ

0.01kΩ

0.5%+2

0.5%+2

0.5%+2

1.0%+3

2.0000MΩ

(2.000MΩ)

0.0001MΩ

1.0%+3

1.0%+3

1.0%+3

0.001MΩ

2.0%+3*

20.000MΩ

(20.00MΩ)

0.001MΩ

2.0%+3*

2.0%+3*

0.01MΩ

2.0%+3*

200.0MΩ

0.1MΩ

2.0%+3*

2.0%+3*

* Do open/short calibration before measuring for above ranges with * to have better precision measurements.

 

DCR

Range

Resolution

Accuracy

200.00Ω

0.01Ω

1.0%+3*

2.0000kΩ

0.0001kΩ

0.2%+2

20.000kΩ

0.001kΩ

0.2%+2

200.00kΩ

0.01kΩ

0.5%+2

2.0000MΩ

0.0001MΩ

1.0%+3

20.000MΩ

0.001MΩ

2.0%+3*

200.0MΩ

0.1MΩ

2.0%+3*

** Do open/short calibration before measuring for above ranges with * to have better precision measurements.

 

Capacitance (Parallel / Series mode):

Range

Resolution

100/120Hz

1kHz

10kHz

100kHz

200.00pF

0.01pF

1.2%+5*

2.0%+5*

2000.0pF

0.1pF

2.0%+3*

0.3%+2

0.6%+3

20.000nF

0.001nF

2.0%+3*

0.3%+2

0.3%+2

0.6%+3

200.00nF

0.01nF

0.3%+2

0.3%+2

0.3%+2

0.6%+3

2000.0nF

0.1nF

0.3%+2

0.3%+2

0.6%+2

2.0%+5*

20.000μF

(20.00μF)

0.001μF

0.3%+2

0.6%+2

1.2%+5*

0.01μF

3.0%+5 (10μF max.)*

200.00μF

(200.0μF)

0.01μF

0.6%+2

1.0%+3*

0.1μF

3.0%+5 (100μF max.)*

2000.0μF

(2000μF)

0.1μF

1.0%+3*

1μF

1.2%+3*

20.00mF

0.01mF

1.2%+3*

*If reading <2000, unit on display is pF

* Do open/short calibration before measuring for above ranges with * to have better precision measurements.

 

Inductance (Parallel / Series mode).

Range

Resolution

100/120Hz

1kHz

10kHz

100kHz

20.000μH

0.001μH

2.5%+5*

200.00μH

0.01μH

1.2%+5*

0.6%+3

2000.0μH

0.1μH

2.0%+5*

0.6%+3

0.6%+3

20.000mH

0.001mH

1.2%+5*

1.0%+5

0.3%+2

0.6%+3

200.00mH

0.01mH

0.3%+2

0.6%+3

0.3%+2

1.2%+5*

2000.0mH

0.1mH

0.3%+2

0.3%+2

0.6%+3

20.000H

0.001H

0.3%+2

0.6%+3

1.2%+5*

200.0H

0.1H

0.6%+3

1.2%+5*

2.000KH

0.001KH

1.2%+5*

*If reading <2000, unit on display is μH

* Do open/short calibration before measuring for above ranges with * to have better precision measurements.

 

Accuracy v.s. Resistance (ZDUT)

 

DCR

100/120Hz

1kHz

10kHz

100kHz

0.1~1Ω

1.2%+5*

1.2%+5*

1.2%+5*

1.2%+5*

2.5%+5*

1~10Ω

0.6%+3*

0.6%+3*

0.6%+3*

0.6%+3*

1.2%+5*

10~100kΩ

0.3%+2

0.3%+2

0.3%+2

0.3%+2

0.6%+3

100k~1MΩ

0.6%+3

0.6%+3

0.6%+3

0.6%+3

2.5%+5*

100k~2MΩ

1M~20MΩ

1.2%+5*

1.2%+5*

1.2%+5*

2.5%+5*

>20MΩ

2.5%+5*

2.5%+5*

2.5%+5*

* Do open/short calibration before measuring for above ranges with * to have better precision measurements.


  TEMP./Humidity:

  • Operation:0°C~ 50°C(32℉~122℉)/<70% R.H.(no condensation)
  • Storage:-20°C~ 60°C(-4℉~140℉)/<80%R.H.(no condensation)

 

  Dimension & Weight:

  • 188(L)x95(W)x52(H)mm
  • Approx.350g (excluding battery)

 

  Accessories:

  • Alligator test lead case (TL-21) x 1
  • AC/DC Adaptor x 1
  • Guard Line (TL-23) x 1
  • DC 9V Battery x1
  • Instruction Manual x1
  • Carry case x 1

 

  Additional Accessories for with USB Function:

  • Optical IR to USB PC link x 1
  • CD Rom (software) x 1
  • TL-22 SMD Tweezers x 1
  • Auto L/C/R check
  • 100 / 120 / 1K / 10K / 100K Hz test frequency
  • 20,000 / 2,000 counts display
  • Backlight
  • Ls/ Lp/ Cs/ Cp/ Rs/ Rp/ DCR with D/Q/θ/ESR measurement
  • Relative mode
  • Series / Parallel modes
  • Components sorting function
    *Selectable tolerance ±0.25%、±0.5%、± 1%、±2%、±5%、±10%、±20%、-20%+80%